In the newer versions of software, you right click on the top of the Workstation toolbar at the top of the screen and select 'enable/disable' modules. You might try checking to see what modules are capable of being configured with workstation. There isn't a separate module for the GC detector. At least in the more recent versions of Workstation we've worked with, only the 3800 GC module should appear in the active/available modules window. On this very old revision of software, it's hard to say whether the system requires a 'module' to be specifically configured for processing/display of your detector signal.
Varian star chromatography workstation manual update#
I agree that a software update could be useful. Front detector status, or oven temp., etc).

The exact message is 'Cannot find MINIMUM.MTH cannot preset STARTUP1.MTH for module 3800' I can control all of the oven/injector/detector parameters from within Star - the FID is lit, the System Control window shows the zone temperatures all at the correct settings.I can even see the signal in the display however, nothing is being plotted in the Display - even though I can select what to display (e.g. Control of the 3800 GC was supported through at least version 6.9.2 of Workstation software.

You should definitely consider upgrading if this is an option. 4.5 is a very old version of Workstation. If so, by now you should be presumably seeing the FID signal in Workstation. Reactivate this new method in Workstation system control and see if it will activate and send the method parameters to your GC. Be sure to set the other parameters of your method accordingly such as oven temperature, column flow, injector temperature/split, etc.
Varian star chromatography workstation manual manual#
You can set the air & hydrogen makeup flows as well in this section, unless you have a FID with manual pneumatics. In the method builder, under the detector section, ensure that the FID electronics are turned on and also that the FID oven is on (probably 250-300C depending upon your application). If you are not seeing the FID signal in Workstation, you most likely don't have it acquiring data. At this point, I would suggest opening the startup.mth method in Workstation method builder. I have never seen another method 'required' for one particular method to activate. Ariel - That is a very peculiar error message. I am tempted to uninstall the software and re-install Star 4.5 from diskette, but am afraid to mess anything up (i have already tried and failed to re-install the 8100 autosampler.).ĭoes anyone know anything about minimum.mth and startup.mth? Thanks, Ariel. So I don't know why the computer is not finding it (i can't find in any of the settings if a directory must be specified). I have searched the computer for minimum.mth, and it is there. When I start a run, the signal does not appear on the plot, and I cannot do any analysis (although on the GC display I can see the FID signal). However, when I load the method, I get an error message that is summarized as: 'cannot find minimum.mth for startup.mth'. The instrument shows up on the Star software, and I can control oven, injector and detector conditions on the GC. CP-3800 GC Getting Started Manual 31 Varian Test.